APPENDIX A - USER CALIBRATIONS
Ohaus Corporation www.ohaus.com
64
Explorer
®
Series Service Manual
A.2 LINEARITY CALIBRATION
The Explorer
®
balances do not have user linearity calibration.
Section 4.4.4 has the procedure to determine if the balance linearity is within specification. See
Appendix B
for the Service Linearization procedures.
A.3 TOUCH SCREEN CALIBRATION
The Touch Screen needs to be re-calibrated if the screen does not respond correctly to intended
commands. A Touch Screen calibration is required if the Touch Screen or Main PCBA is replaced.
Menu
User Settings
Display
Settings
Touch Calibrate
With the balance turned ON, touch
Menu
then
User Settings
then
Display Settings
and then
Touch Calibrate
.
Touch Calibrate:
“Touch the screen at the center of the ring.
O
Press here.”
(First top-left, then bottom-right)