APPENDIX C - SOFTWARE SERVICE TOOL INSTRUCTIONS
Explorer
®
Series Service Manual
73
Ohaus Corporation www.ohaus.com
General Notes:
x
Commands sent to the indicator must be terminated by a carriage return (CR) or a carriage
return-line feed (CRLF).
x
Alternate command characters may be defined by the user.
x
Data output is always terminated with a carriage return-line feed (CRLF).
x
There is 40-second timeout control for print under stable requirement. If the unstable
condition continues over 40 seconds, balance will respond “ES”.
Note 2: Application list
Index
Application
Index
Application
0
Weighing
11
Pipette Adjustment
1
Parts Counting
12
Ingredient Adjustment
2
Percent Weighing
13
SQC
3
Check Weighing
4
Dynamic Weighing
5
Filling
6
Totalization
7
Formulation
8
Differential
9
Peak Hold
10
Density
Determination
Note 3: Unit list
Index
Unit
Index
Unit
0
g
11
dwt
1
kg
12
mo
2
t
13
msg
3
mg
14
HKt
4
15
SGt
5
ct
16
TWt
6
N
17
tical
7
lb
18
tola
8
oz
19
baht
9
ozt
20
10
GN
21
custom_1
22
custom_2
23
custom_3