CHAPTER 3 - MAINTENANCE PROCEDURES
Explorer
®
Series Service Manual
41
Ohaus Corporation www.ohaus.com
Re-install:
Reverse the procedures to re-install the PCBAs. See
Appendix C
for instructions on how to use the
software Service Tool to configure the new PCBA to the balance.
5. Remove
Main PCB (EX12001, EX24001, EX35001)
•
Disconnect the cables along the top edge of the PCB.
•
Loosen the 2 retaining screws on each side of the PCB with approximately 2 turns
•
Rotate the retaining washers so that it no longer engages the PCB, slide out the PCB
Re-install:
Reverse the procedures to re-install the PCBAs. See
Appendix C
for instructions on how to use
the software Service Tool to configure the new PCBA to the balance.
Connect PCB Cable
Power Jack cable
Top edge connectors
Real Time Clock battery