Functional Test
OS 4
VV016042, 2021-03-23 / Rev. 09
3.
Settings in service mode
Pedal (Status / Pedal)
OK
3.1
Heel, side and top switch, each left/right, is correctly displayed.
Notice: _____________________________________________________________________________
3.2
Neutral position: no deflection is displayed (all fields are black)
Notice: _____________________________________________________________________________
3.3
Vertical Pos. 1, Pos. 2, Pos. 3 and Pos. 4: sectors from 0 to 100 % are correctly displayed and correspond
to the mechanical transition.
Notice: _____________________________________________________________________________
3.4
Horizontal left and right: Sectors from 0 to 100 % are correctly displayed.
Notice: _____________________________________________________________________________
3.5
Check cable connection.
Notice: _____________________________________________________________________________
3.6
Check wireless connection.
Notice: _____________________________________________________________________________
Sensors (Calibration / Sensors)
Without cassette!
OK
3.7.
Force sensor
0 ±10
Current value: _________ mmHg
New value: _________ mmHg
Notice: _____________________________________________________________________________
3.8
Venturi sensor
0 ±10
Current value: _________ mmHg
New value: _________ mmHg
Notice: _____________________________________________________________________________
3.9
Control air sensor
0 ±10
Current value: _________ mmHg
New value: _________ mmHg
Notice: _____________________________________________________________________________
3.10
Pressport air sensor
0 ±10
Current value: _________ mmHg
New value: _________ mmHg
Notice: _____________________________________________________________________________
3.11
Injection sensor
0 ±0.05
Current value: _________ bar
New value: _________ bar
Notice: _____________________________________________________________________________
3.12
Extraction sensor
0 ±0.05
Current value: _________ bar
New value: _________ bar
Notice: _____________________________________________________________________________
3.13
VIT PN sensor
0 ±0.05
Current value: _________ bar
New value: _________ bar
Notice: _____________________________________________________________________________
Only calibrate in case of a deviation from tolerance!
Содержание OS 4
Страница 1: ......
Страница 44: ...2021 03 23 Rev 10 VV016042 44 10 2 Core PCB ...
Страница 49: ...2021 03 23 Rev 10 VV016042 49 From To JP100 Speaker right JP200 JP201 JP202 BAT700 Support battery ...
Страница 134: ...2021 03 23 Rev 09 VV016042 134 18 Appendix C Safety check Please see form on next page ...
Страница 136: ...2021 03 23 Rev 09 VV016042 136 19 Appendix D Functional test Please see form on next page ...
Страница 147: ...2021 03 23 Rev 09 VV016042 147 20 Appendix E Laser installation protocol Please see form on next page ...
Страница 151: ...VV016042 2021 03 23 Rev 10 ...