2021-03-23 / Rev. 10
VV016042
17
7.4
Sensor calibration
Calibration
sensors
Store actual offset values of the sensors
displayed by touching <
Calibrate
>.
DANGER!
Incorrect calibration
Risk of eye damage to pa-
tient
► No instruments shall be
connected to the device
while calibrating sensors.
7.5
Pedal calibration
Calibration
pedal
Before adjusting and pairing the
pedal make sure that it is con-
nected with cable to the OS 4.
Turn off the device, wait for 20 sec. at
the minimum
Connect pedal with cable to the OS 4
Turn on the device
Call up the service mode (see
Pedal Adjustment:
The pedal adjustment has to be
carried out before the pedal
pairing is started.
Make sure that the foot pedal is
placed on the ground.
Push <
Start
> to begin with the cali-
bration routine.
Move to pedal position shown on the
sequence and <
Confirm
>.
Pedal Pairing:
In the menu <Calibration> select <Pe-
dal>
Push the button <Start Pairing>
As soon as the pairing is finished, the
message “Success” appears.
Establish radio communication:
Connect pedal with cable to the OS 4
Push the button <Enable Wireless>
Turn off radio communication:
Connect pedal with cable to the OS 4
Push the button <Disable Wireless>
Содержание OS 4
Страница 1: ......
Страница 44: ...2021 03 23 Rev 10 VV016042 44 10 2 Core PCB ...
Страница 49: ...2021 03 23 Rev 10 VV016042 49 From To JP100 Speaker right JP200 JP201 JP202 BAT700 Support battery ...
Страница 134: ...2021 03 23 Rev 09 VV016042 134 18 Appendix C Safety check Please see form on next page ...
Страница 136: ...2021 03 23 Rev 09 VV016042 136 19 Appendix D Functional test Please see form on next page ...
Страница 147: ...2021 03 23 Rev 09 VV016042 147 20 Appendix E Laser installation protocol Please see form on next page ...
Страница 151: ...VV016042 2021 03 23 Rev 10 ...