2021-03-23 / Rev. 10
VV016042
64
11.1.7
Power supply 24 VDC
VX320019
Quantity
Description
1
Power supply unit 24V
STEP 1:
1.
2.
Cables are fixed to the body with ca-
ble ties. Remove them all.
Remove black ring that fixes the
compressor by loosening 4 screws.
The compressor cable is routed
through a ring. Therefore, leave the
ring in the unit.
STEP 2:
3.
4.
5.
6.
Disconnect all cables from the
power supply.
Remove 4 screws from the supply
carrier (sheet metal).
Turn supply carrier over.
Remove 4 screws from the backside
of the carrier and take power supply
away.
For connection scheme of wires, see
NOTE!
Use of incorrect screws
Risk of a short circuit
► Do not use any other screws
than the original ones intended
to fix the power supply.