2021-03-23 / Rev. 10
VV016042
126
14.2
Error Messages
Error number
Meaning
Solution
ERROR 1
Supply error
+12V, +24V, +5V, -5V, +3.3V not within tol-
erance
Replace system unit.
ERROR 2
Adjust DIA
Replace the system unit and notify the manufacturer.
ERROR 3
Pedal not calibrated or incorrectly cali-
brated
Adjust pedal
The pedal has to be readjusted as described in the
Section “Service mode operation / Pedal calibration”.
ERROR 4
Remote graphic display
- Check the cable to the remote control panel.
- Replace the remote control panel.
ERROR 5
Display error
(RESET key is blinking)
- Check the cable from the carrier board PCB to the
control unit.
- Replace the control panel.
ERROR 7
CORE software
Restart device, send Log-Files to manufacturer
ERROR 8
COM error
(RESET key is blinking)
- Restart device, send Log-Files to manufacturer
- Check the cable from the carrier board PCB to the
core PCB
- Replace the carrier board PCB
ERROR 10
Peristaltic pump failure
Check the encoder cable and the motor cable to the
peristaltic pump motor.
ERROR 11
HF output stage
Replace the system unit and notify the manufacturer.
ERROR 12
Irrigation pole motor
- Check if irrigation pole is blocked or overloaded
- Check the cables of the drive for infusion pole
- Replace the drive for infusion pole
ERROR 15
Force sensor
- Adjust force sensor
- Ascertain whether the force sensor is plugged in or
whether there is a severed cable
- Replace fluidic unit.
ERROR 16
Adjust Venturi sensor
Adjust Venturi sensor or replace fluidic PCB.
ERROR 17
Adjust force sensor
Adjust force sensor or replace fluidic unit.
ERROR 18
Check Venturi pump system
- Adjust Venturi and force sensor
- Replace fluidic unit
ERROR 20
Extraction deviation
- Replace compressor unit
- Replace visco module
ERROR 21
Adjust Visco sensor
Adjust injection and extraction sensor.
ERROR 22
Adjust air sensor
Adjust AIR sensor and pressport AIR sensor.
ERROR 23
VIT PN differential sensor
Replace VIT-PN module
ERROR 24
LUM1 module
- Wait for a short time while light module can cool
down.
- Replace LEDplus module, light LEDplus cannot be
operated.
ERROR 25
LUM2 module
- Wait for a short time while light module can cool
down.
- Replace LED module, Light LED cannot be operated.
ERROR 26
Venturi System
- Adjust Venturi Sensor
- Replace Venturi unit
ERROR 31
Liquid level sensor
Notify the manufacturer
ERROR 32
Irr valve
- Replace irrigation valve
- Replace fluidic unit
ERROR 33
Air/gfi port
- Adjust Control AIR sensor and pressport AIR sensor.
- Replace AIR/GFI module
ERROR 38
Laser module
- Switch off device, disconnect it from the mains sup-
ply, connect it to the mains supply, switch device on
- Notify manufacturer
ERROR 39
Wireless module
- Restart device
- Replace wireless PCB
Содержание OS 4
Страница 1: ......
Страница 44: ...2021 03 23 Rev 10 VV016042 44 10 2 Core PCB ...
Страница 49: ...2021 03 23 Rev 10 VV016042 49 From To JP100 Speaker right JP200 JP201 JP202 BAT700 Support battery ...
Страница 134: ...2021 03 23 Rev 09 VV016042 134 18 Appendix C Safety check Please see form on next page ...
Страница 136: ...2021 03 23 Rev 09 VV016042 136 19 Appendix D Functional test Please see form on next page ...
Страница 147: ...2021 03 23 Rev 09 VV016042 147 20 Appendix E Laser installation protocol Please see form on next page ...
Страница 151: ...VV016042 2021 03 23 Rev 10 ...