2021-03-23 / Rev. 10
VV016042
41
9.7.4.1
Calibrate TLD (Target Laser Diode)
Calibration of the TLD is performed on this page. The current power value is shown and the new one is entered here.
NOTE: The entered power value must be 40 or higher in order to achieve that the value of the aiming beam reaches the
specified value and tolerance range.
For GLM01, adapter
<Endo>
is used. Make sure the correct adapter is selected during calibration.
<Read Config>
reads the values from the module and displays them.
After calibration, the configuration must be written on the module by
<Write Config>
.
The configuration can be saved into a file by
<Export Config>
and restored by
<Import Config>
.