2021-03-23 / Rev. 10
VV016042
131
Test fibre
Fibre: 200µm, NA 0.22
Connector: SMA (both ends)
Safety version with protective tube and bend lim-
itation.
Available at:
Oertli Instrumente AG,
Article No.: MHP200L02
Power meter
Recommended:
Ophir Optronics Solutions Ltd, www.ophi-
ropt.com
OPHIR NOVA II
Article No. 7Z01550
Measurement head
Recommended:
Ophir Optronics Solutions Ltd, www.ophi-
ropt.com
Ophir 3A-P-SH-V1
Article No. 7Z02622
SMA fibre adapter for 3A-P-SH-V1
Recommended:
Ophir Optronics Solutions Ltd, www.ophi-
ropt.com
Article No. 1G01236
CAUTION!
The laser power meter must be cali-
brated periodically!