2021-03-23 / Rev. 10
VV016042
10
3.
If necessary, replace the defective fuse.
4.
Insert the voltage selector in such a way that the correct mains voltage is displayed in the viewing window.
- 115 V: 115 V must be displayed in the viewing window of the voltage selector.
- 230 V: 230 V must be displayed in the viewing window of the voltage selector.
5.
Reinsert the fuse drawer.
6.3
Infusion pole
1.
Insert (4) infusion pole (1) into the holder (3).
2.
Firmly rotate the infusion pole in the direction of the arrow (5).
3.
If the infusion hooks are not at a 45-degree angle to the device, proceed as follows:
Use an Allen key to loosen the screw (2), position infusion pole at a 45-degree angle as illustrated above and secure
screw once more.
6.4
Door remote switch (DRS) for endo laser module
1.
Unscrew the cap (1).
2.
Remove the cover (2).
3.
Remove the cable bridge (3). Use these pins to connect the door switch.