TUF-2000F User Manual
25
0056
0221-0222
2
Inner pipe diameter
REAL4
In millimeter
0229-0230 2
Upstream
delay
REAL4 In
microsecond
0231-0232
2
Downstream delay
REAL4
In microsecond
0233-0234 2
Calculated
travel
time REAL4
In
microsecond
0257-0288
32
LCD buffer
BCD
0289
1
LCD buffer pointer
INTEGER
0311
2
Worked time for today
LONG
Unsigned, in
seconds
0313
2
Worked time for this month
LONG
Unsigned, in
seconds
1437
1
Unit for flow rate
INTEGER
See note 5
1438
1
Unit for flow totalizer
INTEGER
Range 0~7,see
note 1
1439
1
Multiplier for totalizer
INTEGER
Range 0~7,see
note 1
1440
1
Multiplier for energy
accumulator
INTEGER Range
0~10,see
note 1
1441
1
Unit for energy rate
INTEGER
0=GJ 1=Kcal
2=KWh
,
3=BTU
1442
1
Device address
INTEGER
1451
2
User scale factor
REAL4
1521
2
Manufacturer scale factor
REAL4
Read only
1529
2
Electronic serial number
BCD
High byte first
Note
:
(1) The internal accumulator is been presented by a LONG number for the integer part
together with a REAL number for the decimal fraction. In general uses, only the integer part
needs to be read. Reading the fraction can be omitted. The final accumulator result has a
relation with unit and multiplier. Assume N stands for the integer part (for the positive
accumulator, the integer part is the content of REG 0009, 0010, a 32-bits signed LONG
integer,), Nf stands for the decimal fraction part (for the positive accumulator, the fraction part
is the content of REG 0011, 0012, a 32-bits REAL float number,), n stands for the flow
multiplier (REG 1439).
then
The final positive flow rate=(N+Nf ) ×10n-3 (in unit decided by REG 1438
)。
The meaning of REG 1438 which has a range of 0~7 is as following:
0 cubic meter (m
3
)
1 liter (L)
2 American gallon (GAL)