TUF-2000F User Manual
11
20. Timed alarm #2 21.Batch Totalizer Full
22. Periodically M51 Timer 23. Oct Not Using
The OCT circuit does not source voltage at its output. It must be connected with an
external power and pull-up resistant for some occasions.
When the OCT circuit is close, it will draw current. The maximum current shall not
be over 100mA.
Attention: the maximum voltage applied to OCT can not be over 80 volts.
M79
Relay or OCT2 setup
By selecting a proper input source, the RELAY will close when the trigger event
occurs
The available trigger sources are:
0. No Signal 1. Poor Signal
2. Not Ready(No*R) 3. Reverse Flow 4. AO Over 100%
5. FO Over 120% 6. Alarm #1 7. Reverse Alarm #2
8. Batch Control 9. POS Int Pulse 10.NEG Int Pulse
11.NET Int Pulse 12.Energy POS Pulse 13.Energy NEG Pulse
14.Energy NET Pulse 15.MediaVel=>Thresh 16.MediaVelo<Thresh
17.ON/OFF viaRS232 18. Timer (M51 Daily) 19.Timed alarm #1
20. Timed alarm #2 21.Batch Totalizer Full
22. Periodically M51 Timer 23. Disable Relay
The RELAY is of SPST (Single pole, single throw) type. It is rated for 110VAC max
and have a current rating of 0.5A resistive load.
It highly recommended that a salve relay to be utilized whenever a large resistive
load or inductive load is to be controlled.
Note. In order to make the user interface compatible with the former version7, the
name RELAY was used other than OCT2, but in fact it is an OCT output.
M80
Window for selecting the trig signal for the built-in batch controller. Available trig
sources:
0. Key input (press ENT key to start the batch controller)
1. Serial port
2. AI3 rising edge (when AI3 receives 2mA or more current)
3. AI3 falling edge (when AI3 stop receiving 2mA or more current)
4. AI4 rising edge (when AI3 receives 2mA or more current)
5. AI4 falling edge (when AI3 stop receiving 2mA or more current)
6. AI5 rising edge (when AI3 receives 2mA or more current)
7. AI5 falling edge (when AI3 stop receiving 2mA or more current)
8.Timer periodically (define the start time and interval time in M51)
9.Timer daily (define the start time and interval time in M51)
For the input analog current signal, 0 mA indicates “0”, 4mA or more indicates ‘1’.
By selecting item #8, the batch totalizer can be started periodically by the internal
timer located at Menu51. When the batch totalizer is full, a signal which indicate the
batch is full can be direct to either the OCT or the RELAY terminals to stop the
pump or other devices.
By selecting item #9, the batch totalizer could act as totalizer witch runs for only a
period of the day so that a alarm signal could be produced if the total flow during
that time period is over a certain amount of. For example, if you want a alarm signal