Operation
OPM223/253
56
5.4.7
Service
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
S
SERVICE function
group
Service function settings.
S1
Select language
ENG = English
GER = German
FRA = French
ITA = Italian
NL = Dutch
ESP = Spanish
This field has to be configured once
during device configuration. Then you
can exit S1 and continue.
S2
Configure Hold
S+C = Hold during
configuration and
calibration
Cal = Hold during calibration
Setup = Hold during
configuration
None = no Hold
S = setup
C = calibration
S3
Manual Hold
Off
On
The setting is retained even in the event
of a power failure.
S4
Enter Hold dwell
period
10 s
0 ... 999 s
S5
Enter SW upgrade
release code (Plus
Package)
0000
0000 ... 9999
If an incorrect code is entered, you are
taken back to the measurement menu.
The number is edited with the PLUS or
MINUS key and confirmed with the
ENTER key.
"1" is displayed if the code is active.
S6
Enter SW upgrade
release code
Chemoclean
0000
0000 ... 9999
If an incorrect code is entered, you are
taken back to the measurement menu.
The number is edited with the PLUS or
MINUS key and confirmed with the
ENTER key.
"1" is displayed if the code is active.
S7
Order number is
displayed
If the device is upgraded, the order code
is automatically adjusted.
S8
Serial number is
displayed
S
SERVICE
S1
Language
ENG
S2
Auto HOLD
S+C
S3
Man
.
HOLD
off
S4
s
Cont
.
Time
10
S5
0000
PlusCode
S6
CleanCode
0000
S7
PR0005
order
S8
12345678
SerNo