OPM223/253
Operation
61
C17
Zero point (zero
point / U_is) is
displayed
Glass:
pH 7.00
pH 5.00 ... 9.00
Antimony:
pH 1.00
pH -1.00 ... 3.00
ISFET:
current value
-500 ... +500 mV
For ISFET, the zero point is displayed in
mV.
C18
Calibration status is
displayed
o.k.
E xxx
C19
Store calibration
result?
yes
no
New
If C18 = E xx, then only No or
New
.
If New, return to C.
If Yes/No, return to "Measurement".
The electrode can now be reinstalled in the process.
C (2)
CALIBRATION
function group:
Calibration for ORP
mV
Calibration ORP mV
Only for A1 = ORP (mV).
The measuring transmitter has a calibrated mV display range. An
absolute mV value with a single buffer solution (adjustment of
measuring chain offset) is set. Here, a buffer solution is used,
preferably with 225 or 475 mV.
The maximum permitted calibration
offset is ±100 mV.
C21
Enter the mV value
belonging to the
ORP buffer used
Current measured value
1500 ... 1500 mV
In the case of symmetrical
measuring operation, the potential
matching pin must also be immersed
in the buffer.
C22
Calibration is
performed
mV value
Stability check:
The value is accepted in the event of
stability
≤
±1 mV for more than 10 s.
C23
Zero point is
displayed
-100 ... 100 mV
C24
Calibration status is
displayed
o.k.
E xxx
C25
Store calibration
result?
yes
no
new
If C24 = E xxx, then only No or
New
.
If New, return to C.
If Yes/No, return to "Measurement".
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
C17
pH
Zero
7
.
00
C18
Status
o
.
k
.
C19
Store
yes
C
CALIBRAT
C21
mV
mV Buffer
225
C22
mV
CAL mV
225
C23
mV
Zero
0
C24
Status
o
.
k
.
C25
Store
yes