Operation
OPM223/253
50
R211
Switch function of
R2 (1) off or on
Off
On
All the settings are retained.
R212
Enter the switch-on
point of the contact
pH 16.00
pH -2.00 ... 16.00
1500 mV
-1500 ... 1500 mV
100.0 %
0.0 ... 100.0 %
Never set the switch-on point and the
switch-off point to the same value!
(Only the operating mode selected in A1
is displayed.)
R213
Enter the switch-off
point of the contact
pH 16.00
pH -2.00 ... 16.00
1500 mV
-1500 mV ... 1500 mV
100.0 %
0.0 ... 100.0 %
Entering a switch-off point selects either
a Max contact (switch-off point < switch-
on point) or a Min contact (switch-off
point > switch-on point), thereby
implementing a hysteresis that is
constantly required (see "Illustration of
the alarm and limit functions" figure).
R214
Enter pick-up delay
0 s
0 ... 2000 s
R215
Enter drop-out
delay
0 s
0 ... 2000 s
R216
Enter alarm
threshold
pH 16.00
pH -2.00... 16.00
1500 mV
-1500 ... 1500 mV
100.0 %
0.0 ... 100.0 %
If the alarm threshold is undershot/
overshot, this triggers an alarm with the
error message and error current at the
transmitter (note alarm delay in field F3).
If defined as a Min contact, the alarm
threshold must be < switch-off point.
R217
Display status for
limit contactor
MAX
MIN
Display only.
R2 (2)
Configure limit
contactor for
temperature
measurement
LC PV = limit contactor pH/
ORP (1)
LC C = limit contactor T
(2)
PID controller (3)
Timer (4)
Clean = Chemoclean (5)
Neutra controller
By confirming with ENTER, another
relay function already switched on is
switched off and its settings are reset to
the factory settings.
R221
Switch function of
R2 (2) off or on
Off
On
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
R211
Function
off
pH
R212
On value
16
.
00
pH
R213
Off value
16
.
00
s
R214
On Delay
0
s
R215
Off Delay
0
pH
R216
A
.
Thresh
16
.
00
R217
LC State
MAX
R2
Sel
.
Type
LC °C
R221