OPM223/253
Trouble-shooting
79
E041
Calculation of calibration
parameters aborted
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Yes
No
—
—
1
E042
Distance of calibration value for
buffer pH 2 from zero (pH 7) too
short
Use buffer solution that is at least
Δ
pH = 2
from electrode zero point.
Yes
No
—
—
1
E043
Distance between calibration
values for pH 1 and pH 2 too
short
Use buffer solutions that are at least
Δ
pH = 2
apart.
Yes
No
—
—
1
E044
Stability requirement for
calibration not fulfilled
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Yes
No
—
—
1
E045
Calibration aborted
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Yes
No
—
—
1
E055
Below main parameter
measuring range
Check measurement and connections; check
device and measuring cable with simulator if
necessary.
Yes
No
No
E057
Main parameter measuring
range exceeded
Yes
No
No
E059
Below temperature measuring
range
Yes
No
No
E061
Temperature measuring range
exceeded
Yes
No
No
E063
Below current output range 1
Check configuration in the "Current outputs"
menu; check measurement and connections;
check device and measuring cable with
simulator if necessary.
Yes
No
No
E064
Current output range 1 exceeded
Yes
No
No
E065
Below current output range 2
Yes
No
No
E066
Current output range 2 exceeded
Yes
No
No
E067
Set point exceeded controller 1
Check configuration.
Yes
No
No
E068
Set point exceeded controller 2
Yes
No
No
E069
Set point exceeded controller 3
Yes
No
No
E070
Set point exceeded controller 4
Yes
No
No
E080
Current output 1 range too small
Increase range in "Current outputs" menu.
No
No
—
—
1
E081
Current output 2 range too small
No
No
—
—
1
E085
Incorrect setting for error current
If the current range "0 ... 20 mA" was selected
in field O311, the error current "2.4 mA" may
not be set.
?
?
?
?
E100
Current simulation active
No
No
—
—
1
E101
Service function yes
Switch off service function or switch device off
and then on again.
No
No
—
—
1
E102
Manual mode active
No
No
—
—
1
E106
Download yes
Wait for download to finish.
No
No
—
—
1
E116
Download error
Repeat download.
No
No
—
—
1
E147
Sensor communication faulty
Check that the sensor is correctly connected,
the cable ends are correctly wired at the
terminals and the cable is not damaged.
No
No
No
E152
PCS alarm
Check sensor and connection.
No
No
No
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom.
cleaning start
Facty User
Facty User
Facty User