Wiring
OPM223/253
18
"
Caution!
• Terminals marked NC may not be wired.
• Unmarked terminals may not be wired.
!
Note!
Please label the sensor terminal block with the sticker provided.
4.1.2
Measuring cable and sensor connection
You require screened special measuring cables to connect pH and ORP electrodes to the
transmitter. The following multi-core and ready-to-use cable types can be used:
Structure and termination of the measuring cables
C07-OPM2x3xx-04-06-00-en-004.eps
Fig. 12: Structure of the special measuring cables
!
Note!
For further information on the cables and junction boxes, please refer to the "Accessories"
section.
Sensor type
Cable
Extension
Electrode without temperature sensor
OPK1
VBA / VBM box + OYK71 cable
Electrode with temperature sensor Pt 100
and TOP 68 plug-in head
OPK9
VBA / VBM box + OYK71 cable
ISFET sensor with temperature sensor
Pt 100 / Pt 1000 and TOP 68 plug-in head
OPK12
VBA / VBM box + OYK12 cable
pH individual electrode with separate
reference electrode and separate
temperature sensor
OPK2
VBA / VBM box + PMK cable
OPK1
OPK2
OPK9
OPK12
PM
pH / Redox
PM
pH or Redox
Ref. or Redox
Pt 100
PM
pH + Pt 100 / Pt 1000
PM
pH + Pt 100 / Pt 1000
Coax Ref.
pH or Redox
BN PM
BN PM
pH or Redox
RD Redox or Ref.
BU Ref.
GN Temp.
YE Temp.
WH Temp.
Coax BK Ref.
inner pH
GN Temp.
WH Temp.
YE Temp.
BN PM
Coax BK
inner
GN Temp.
WH Temp.
YE Temp.
RD DRN
BN PM
Ref
SRC IsFET
Содержание OPM223
Страница 2: ......
Страница 95: ...OPM223 253 Technical data 95...
Страница 100: ...Appendix OPM223 253 100...
Страница 103: ...OPM223 253 103...
Страница 104: ...BA194e 00 07 09 71106576 Printed in Germany FM SGML 6 0 DT 71106576...