OPM223/253
Operation
39
5.4.5
Monitoring functions
You can use the monitoring functions to define various alarms and configure output contacts.
Each individual error can defined to be effective or not (at the contact or as an error current).
Moreover, the electrode can be checked for glass breakage or leak current (P1, P2, P7). In the
event of an alarm, a cleaning function can also be activated (F8).
!
Note!
You can only check for glass breakage or leak current and avail of the cleaning function in the
Plus package.
Alarm
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
F
ALARM function
group
Alarm function settings.
F1
Select contact type
Latch = latching contact
Momen = momentary contact
The contact type selected only applies to
the alarm contact.
F2
Select time unit
s
min
F3
Enter alarm delay
0 s (min)
0 ... 2000 s (min)
Depending on the option selected in F2,
the alarm delay is entered in s or min.
F4
Select error current
22 mA
2.4 mA
This selection must be made even if all
error reporting is switched off in F5.
"
Caution!
If "0-20 mA" was selected in O311, "2.4
mA" may not be used.
F5
Select error
1
1 ... 255
Here you can select all the errors which
should trigger an alarm.
The errors are selected via the error
numbers. Please refer to the table in
section "System error messages" for the
meaning of the individual error numbers.
The factory settings remain in effect for
all errors not edited.
F6
Set alarm contact to
be effective for the
selected error
yes
no
If “no” is selected, all the other alarm
settings are deactivated (e.g. alarm
delay). The settings themselves are
retained.
This setting
only
applies to the error
selected in F5.
F
ALARM
F1
Cont
.
Type
Latch
F2
Time Unit
s
F3
s
Err.Delay
0
F4
Err
.
Curr
22mA
F5
Sel
.
error
1
F6
Rel
.
Assg
yes