OPM223/253
Operation
57
5.4.8
Service 2
S9
Reset the device to
the basic settings
No
Sens = sensor data
Facty = factory settings
Sens = last calibration is deleted and is
reset to factory setting.
Facty = all data (apart from A1 a. S1) are
deleted and reset to the factory setting!
S10
Perform device test
No
Displ = display test
S11
Reference voltage
is displayed
Current value in mV
This is used to check the reference
potential. Value > 50 mV indicates
galvanic voltage in the medium.
High values (> 1000 mV) may falsify the
measured value.
S12
Select AC
frequency
50 Hz
60 Hz
Only select 60 Hz if the frequency of the
voltage at the place of use is 60 Hz, the
measured value is fluctuating or sporadic
SCS errors are reported.
Coding
Field
Setting range
(Factory settings, bold)
Display
Info
S9
S
.
Default
no
S10
Test
no
S11
mV
uncompRef
25
S12
Hz
AC Freq
50
Coding
Field
Setting range
(Factory settings, bold)
Display
Note
E
SERVICE 2
function group
Information on the device version
E1
Select module
Contr = controller
(1)
Trans = transmitter (2)
Main = power unit (3)
Rel = relay module (4)
E111
E121
E131
E141
E151
Software version is
displayed
If E1 = contr: instrument software
If E1 = trans, main, rel: module firmware
E112
E122
E132
E142
E152
Hardware version is
displayed
Only display function
E
SERVICE 2
E1
Select
Contr
E111
SW-Vers
.
xx
.
xx
E112
HW-Vers
.
xx
.
xx