RD 160.iCare+.OPM.ENG.V16 04.27.2020
p. 5
ENG
CHAPTER 4 – INSTALLATION AND SET-UP
Warning
The use of controls, settings or the implementation of procedures other than those specified below
may expose users and patients to danger.
4.1 Unpacking
Warning
Information
Upon receipt of the unit, look for any damage that may have occurred during transportation.
•
When unpacking, check the items described section 3.1 are supplied
•
If necessary, contact your supplier.
•
Keep the original packaging for further use.
•
Keep the packaging away from children.
4.2 Location
Warning
Information
Before installing iCare+, choose a suitable location as described below:
•
The device should be positioned on a flat, solid and level surface
•
The surface must be able to support the weight of the device in conditions of use: 16 kg
•
The device should not be placed near a sink or any source that may cause moisture ingress
•
The device should be placed in a properly ventilated room
•
The device should be kept away from any source of heat
4.3 Installation
4.3.1 Connection to the air system
Insert the air supply tube in the air filter inlet, fixed on the Rear panel of the device: Fig.5. Be sure that the air tube
is attached properly to the filter (see picture below). Connect the common port of the "T-junction" connector to the
air compressor output and connect the output ports to your dental unit and iCare+: Fig.6.
Warning
Information
•
Following the norm EN ISO 7494-2, the air quality provided to iCare+ has to be dry, clean and
free from bacteria and contamination
•
It is necessary to follow the above-mentioned parameters in order to avoid malfunction or damage
to the unit
•
The pressure of the air supply has to be between 5 and 6 bar
•
The minimum air flow level required is 50 L/min
Fig.5 – Air Inlet
Fig.6 – Air connection
Dental Unit
i-care+
T junction
Air Compressor
Connect output of the
air compressor to the
“T” junction, and
outputs of the “T” to
i-Care+ and dental unit