RD 160.iCare+.OPM.ENG.V16 04.27.2020
p. 11
ENG
Caution
Before using iCare+:
•
Please make sure to remove any bur from the chuck
•
Make sure also to remove any support/holder for snap-on caps
•
Failure to do so could result in the dirt and liquid inside the instrument not being expelled
5.3 Removal of the instruments
Once the cycle has finished, you will hear a "Beep" sound and the green operating light indicator "end" will be
illuminated.
•
Open the door
•
Remove the instruments from their couplings as below.
5.3.1 Removal of the turbines
•
To remove a standard turbine with Midwest coupling:
-
Unscrew the retaining nut
-
Remove the turbine by pulling down the instrument
•
Removing a turbine with NSK
®
connection from its adaptor:
-
Move up the retaining ring of the adaptor.
-
Remove the turbine by pulling it down while keeping the retaining ring in the upright position.
•
Removing a turbine with Kavo and Sirona
®
connection from its adaptor:
-
Remove simply the turbine by pulling down on the instrument.
•
Removing a with W&H
®
connection from its adaptor:
-
Move the retaining ring of the adaptor up.
-
Remove the turbine by pulling it down while keeping the retaining ring in the upright position.
•
Removing a turbine with Bien-Air
®
connection with from its adaptor:
-
Press the button on the adapter.
-
Remove the turbine by pulling it down while keeping the button pressed.
Fig.19 – NSK
Fig.20 – Kavo / Sirona
Fig.21 – W&H
Fig.22 – Bien Air
5.3.2 Removal of the contra-angles
For removing a contra-angle, in accordance with the ISO 3964 standard, follow the description below:
•
Start by opening the door
•
Push the release button
•
Remove the contra-angle by pulling it down while the release button is pressed.