“W4” drawer indicator informs the operator about either two kind anomalies:
-
The drawer is not properly positioned or absent. In this case, check if the drawer is present and
correctly inserted. If necessary, extract the drawer and insert it again.
-
The drawer is full and might overflow. In this case, remove the drawer and empty it as described
section 6.2.3.
•
“W5” pressure indicator informs the operator about air pressure problems such low air pressure, air leakage, no
air and excessive air pressure. In such a case, iCare+ cannot restart the cycle until the problem is solved
-
Check the O-ring of the e-type, if necessary, insert new O-rings: Refer to section 6.2.6.
-
Check the air compressor and air connection: Refer to section 4.3.1
•
“W6” door indicator means the door is open or not properly closed. Make sure that nothing obstructs door
closing, and close the door to start any cycle.
The iCare+ is controlled by a micro-processor which checks, in real time, operating conditions. If a
problem occurs, it is detected, and then the iCare+ will stop immediately.
CHAPTER 8: TRACEABILITY
The reprocessing of the instruments has to be performed in accordance with the regulation in force as well as
following the latest recommendations. To ensure the efficiency of the reprocessing performed by iCare+ and
enabling the traceability of the handpieces, reprocessing data is recorded internally and uploaded to the USB key,
refer to section 3.3, Fig.4.
To ensure traceability, the software “S” supplied with the device enables you to check the data relating to
the reprocessing cycles. Please refer to software user manual to retrieve cycle information.
CHAPTER 9: TROUBLE SHOOTING
Problem
Possible cause(s)
Solution(s)
The device does not
switch on
Is the power supply cord connected to
mains?
Connect the power supply cord to mains
Is the power supply cord properly
connected to the iCare+?
Insert the power cord plug into the
device's main connector
Is the main switch positioned in the "ON"
position?
Turn the main switch to the "ON" position
Are the fuses working?
Change the fuses with the same model
The device does not
start the reprocessing
cycle
Is the door open? (door LED illuminated)
Close the door properly
Are the n.clean or n.cid indicators
illuminated?
Change bottles and prime
Is the Oil indicator illuminated?
Fill the Oil tank
The device does not
operate correctly
Is the air supply tube properly connected
to the device?
Connect the air supply tube to the device's
connector
Air pressure supplied between 5 and 6
bar?
Set the air supply pressure between 5 and
6 bar
Is short mode selected?
Select Long mode
Are the instruments properly installed?
Reconnect the instruments securely