NSK iCare+ Скачать руководство пользователя страница 21

 

 

RD 160.iCare+.OPM.ENG.V16 04.27.2020 

 

                 p. 19

 

ENG 

 

“W4” drawer indicator informs the operator about either two kind anomalies: 

-

 

The  drawer is  not  properly  positioned  or  absent.  In  this  case,  check  if  the  drawer  is  present  and 

correctly inserted. If necessary, extract the drawer and insert it again. 

-

 

The drawer is full and might overflow. In this case, remove the drawer and empty it as described 

section 6.2.3. 

 

“W5” pressure indicator informs the operator about air pressure problems such low air pressure, air leakage, no 
air and excessive air pressure. In such a case, iCare+ cannot restart the cycle until the problem is solved 

-

 

Check the O-ring of the e-type, if necessary, insert new O-rings: Refer to section 6.2.6. 

-

 

Check the air compressor and air connection: Refer to section 4.3.1 

 

“W6”  door  indicator  means  the  door  is  open  or  not  properly  closed.  Make  sure  that  nothing  obstructs  door 
closing, and close the door to start any cycle. 

 

The iCare+ is controlled by a micro-processor which checks, in real time, operating conditions.  If a 
problem occurs, it is detected, and then the iCare+ will stop immediately. 

CHAPTER 8: TRACEABILITY 

The  reprocessing  of  the  instruments  has  to  be  performed  in  accordance  with  the  regulation  in  force  as  well  as 
following  the  latest  recommendations.  To  ensure  the  efficiency  of  the  reprocessing  performed  by  iCare+  and 
enabling the traceability of the handpieces, reprocessing data is recorded internally and uploaded to the USB key, 
refer to section 3.3, Fig.4. 

To ensure traceability, the software “S” supplied with the device enables you to check the data relating to 
the reprocessing cycles. Please refer to software user manual to retrieve cycle information. 
 
 

CHAPTER 9: TROUBLE SHOOTING 

Problem 

Possible cause(s) 

Solution(s) 

The device does not 
switch on  

Is the power supply cord connected to 
mains? 

Connect the power supply cord to mains 

Is the power supply cord properly 
connected to the iCare+? 

Insert the power cord plug into the 
device's main connector 

Is the main switch positioned in the "ON" 
position? 

Turn the main switch to the "ON" position 

Are the fuses working? 

Change the fuses with the same model 

The device does not 
start the reprocessing 
cycle 

Is the door open? (door LED illuminated) 

Close the door properly 

Are the n.clean or n.cid indicators 
illuminated? 

Change bottles and prime 

Is the Oil indicator illuminated? 

Fill the Oil tank 

The device does not 
operate correctly 

Is the air supply tube properly connected 
to the device? 

Connect the air supply tube to the device's 
connector  

Air pressure supplied between 5 and 6 
bar? 

Set the air supply pressure between 5 and 
6 bar 

Is short mode selected?  

Select Long mode 

Are the instruments properly installed?  

Reconnect the instruments securely 

Содержание iCare+

Страница 1: ...RD 160 iCare OPM ENG V16 04 27 2020...

Страница 2: ...ion of dynamic dental instruments iCare allows the treatment of the inner and outer surfaces of different instruments contra angles turbines etc Complete treatment in less than 17 minutes Up to 4 inst...

Страница 3: ...njury or damage to the device if the safety instructions are not correctly followed Information Text preceded by this symbol contains useful information for using your iCare device 2 2 General safety...

Страница 4: ...mage the device 2 3 Safety labels Adhesive safety labels have been placed on the iCare in important locations These include storage instructions for the bottles the device and information about the de...

Страница 5: ...is a smell of burning switch off the iCare immediately and contact your retailer Do not use the iCare with flammable gases CHAPTER 3 DESCRIPTION OF THE DEVICE 3 1 Items requested for operation Picture...

Страница 6: ...n cid bottle E Air filter Pressure regulator J USB port for recording data Fig 1 Front view Fig 2 Rear view Fig 3 Operating fluids Fig 4 USB port 3 3 Description of the front panel Table 4 1 Start St...

Страница 7: ...ause moisture ingress The device should be placed in a properly ventilated room The device should be kept away from any source of heat 4 3 Installation 4 3 1 Connection to the air system Insert the ai...

Страница 8: ...checked by a certified electrician A fluctuation in the mains voltage or an electromagnetic field which does not comply with the standard may disrupt the operation of the device 4 4 Consumables 4 4 1...

Страница 9: ...with n cid rinse cautiously with water for several minutes Remove contact lenses if present and easy to do Continue rinsing If eye irritation persists seek medical advice attention In case of contact...

Страница 10: ...dation performed by an unauthorized person could result in your iCare being unsafe for you and your patients Periodical validation is limited exclusively to professional dental installers who have bee...

Страница 11: ...disinfecting solution Once the onsite preparation of the handpieces is complete they will should be safely transported to the reprocessing site avoiding injuries contamination and damage 5 1 2 Preclea...

Страница 12: ...Please refer chapter 10 for the list of available coupling adaptors Fig 14 Connection of a turbine Fig 15 Screwing the retaining nut Fig 16 Coupling adaptor It is necessary to use specific iCare air t...

Страница 13: ...aining ring of the adaptor Remove the turbine by pulling it down while keeping the retaining ring in the upright position Removing a turbine with Kavo and Sirona connection from its adaptor Remove sim...

Страница 14: ...ion Lubrication Partial cycle Cleaning Lubrication Short cycle Lubrication only K7 Spray key In conjunction with K2 to K5 this key is used to define if the handpiece has an external spray channel or n...

Страница 15: ...op corresponds to a normal volume Short cycle Large drop corresponds to a specific volume for certain contra angles long cycle for surgical or 1 5 contra angle for example Unlit light indicators 3 ind...

Страница 16: ...external spray instrument but in reality the instrument has an internal spray An instrument that has been installed with external spray without using the button ext W O An instrument which has been in...

Страница 17: ...urther steps Fig 30 Indicators during cycle in progress Caution When the oil drop indicators are switched off the related instruments are not selected Before first use or after a long period of no use...

Страница 18: ...l panel and never use aggressive detergents 6 2 General maintenance guidelines Frequency Number of cycles Operation Reference Section Every week 50 Cleaning the treatment chamber 6 2 1 Every week 50 C...

Страница 19: ...n and n cid solutions are formulated to be disposed of in the sewage disposal Prior to doing so the soiled fluids should be diluted using water in a 4 1 ratio 4 volume of water for 1 volume of soiled...

Страница 20: ...er use other fuses than those described below Voltage 250VAC Rating T 1 6AH 250V CHAPTER 7 INFORMATION FROM WARNING INDICATORS As shown in Fig 34 warning indicators inform in real time about abnormal...

Страница 21: ...ons To ensure the efficiency of the reprocessing performed by iCare and enabling the traceability of the handpieces reprocessing data is recorded internally and uploaded to the USB key refer to sectio...

Страница 22: ...luminated and a Beep signal can be heard the air pressure is lower than 5 bar or higher than 6 bar To restart the unit first check the air pressure at the inlet of the machine It is necessary to have...

Страница 23: ...bar and 6 bar 5 m Not significant IEC 61010 1 2016 EN 61326 1 2012 ISO 15883 1 ISO 15883 5 ISO 10993 5 Class II b class in accordance with rule 15 of annex IX NSK Europe GmbH Elly Beinhorn Strasse 8 D...

Страница 24: ...ectronic equipment WEEE Directive 2002 96 EC for product and accessory disposal Consult operating instructions Caution refer to attached instructions For indoor use only Use by date Number of items in...

Страница 25: ...012 iCare WH Adaptor Not supplied with the device Must be ordered separately Z1127013 iCare BA Adaptor Not supplied with the device Must be ordered separately Z1127014 4 6 mm diameter tube Supplied wi...

Страница 26: ...r quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 0 Ut during 10ms 0 Ut during...

Страница 27: ...iCare iCare is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of iCare can help prevent electromagnetic interference by maintaining a mini...

Страница 28: ...al Spain SA C M dena 43 El Soho Eur polis 28232 Las Rozas Madrid Spain Tel 34 91 626 61 28 Fax 34 91 626 61 32 info nsk spain es http www nsk spain es NSK United Kingdom Office 4 Gateway 1000 Arlingto...

Отзывы: