NSK iCare+ Скачать руководство пользователя страница 17

 

 

RD 160.iCare+.OPM.ENG.V16 04.27.2020 

 

                 p. 15

 

ENG 

5.5 Launching the reprocessing cycle 

Once the reprocessing cycle is selected and instruments setting achieved, press the “Start Key” (K1) to launch the 
cycle (see Figure 23). 

 

 

“Start Key” launches the selected cycle for the selected instruments. If a problem occurs or if you 
wish to stop the cycle the “Start Key” stops the unit when pushed and held 

 

The cycle is then interrupted, the settings and selections are recorded and the device is in standby 
mode. Pressing “Start Key” again, allows the reprocessing cycle relaunch from the beginning with 
the same settings 

 

In case the door is open while the unit is in standby mode, the settings and selections are then reset 
to default 

 
When reprocessing cycle is launched, the status indicators on the control panel indicate the progress of the cycle 
and the treatment chamber is illuminated using the same colour code. 

 

Yellow indicator = Cleaning phase 

 

White indicator = Disinfection phase 

 

Blue indicator = Lubrication phase 

 

Once the cycle is completed the green indicator is illuminated to inform the user that the instruments are ready 
for further steps. 

 

 

Fig.30– Indicators during cycle in progress  

 

 

Caution

 

 

When the “oil drop” indicators are switched off, the related instruments are not selected 

 

Before first use, or after a long period of no use, run the device without instruments 

 

 

 

 

 

 

 

Caution

 

 

In order to facilitate external cleaning, you need to perform pre-cleaning as defined in section 5.1 

 

Use disinfectant wipes to handle the instruments while connecting them to the unit  

 

At the end of the cycle, open the door and remove the instrument using a disinfectant wipe 

 

Lubrication phase

 

Lubrication is a maintenance 
step. 

Lubricants 

reduce 

noise,  friction  and  prevent 
wear  and  accumulation  of 
wear debris. This step uses a 
high pressure air blowing oil 
inside the instrument.

 

Disinfection phase

 

Disinfection  is  the  reduction 
of  the  number  of  viable 
organisms of the instruments. 
The  devices  operates  using 
chemical  disinfecting  agents 
which  are  used  to  neutralize 
the remaining agents.

 

Purge and Initialisation

 

High  pressure  air  is  used  for 
removal  of  elements  which 
soil  the  instrument  and  clog 
the  internal  channels.  For 
example :  oil  residue,  body 
fluids, and metal particles.

 

Cleaning phase

 

Cleaning  is  the  removal  of 
initial  contamination.  The 
purpose  of  cleaning  is  to 
leave  as  little  residue  as 
possible,  since  this  can 
adversely affect disinfection

 

Operating cycle

 

Содержание iCare+

Страница 1: ...RD 160 iCare OPM ENG V16 04 27 2020...

Страница 2: ...ion of dynamic dental instruments iCare allows the treatment of the inner and outer surfaces of different instruments contra angles turbines etc Complete treatment in less than 17 minutes Up to 4 inst...

Страница 3: ...njury or damage to the device if the safety instructions are not correctly followed Information Text preceded by this symbol contains useful information for using your iCare device 2 2 General safety...

Страница 4: ...mage the device 2 3 Safety labels Adhesive safety labels have been placed on the iCare in important locations These include storage instructions for the bottles the device and information about the de...

Страница 5: ...is a smell of burning switch off the iCare immediately and contact your retailer Do not use the iCare with flammable gases CHAPTER 3 DESCRIPTION OF THE DEVICE 3 1 Items requested for operation Picture...

Страница 6: ...n cid bottle E Air filter Pressure regulator J USB port for recording data Fig 1 Front view Fig 2 Rear view Fig 3 Operating fluids Fig 4 USB port 3 3 Description of the front panel Table 4 1 Start St...

Страница 7: ...ause moisture ingress The device should be placed in a properly ventilated room The device should be kept away from any source of heat 4 3 Installation 4 3 1 Connection to the air system Insert the ai...

Страница 8: ...checked by a certified electrician A fluctuation in the mains voltage or an electromagnetic field which does not comply with the standard may disrupt the operation of the device 4 4 Consumables 4 4 1...

Страница 9: ...with n cid rinse cautiously with water for several minutes Remove contact lenses if present and easy to do Continue rinsing If eye irritation persists seek medical advice attention In case of contact...

Страница 10: ...dation performed by an unauthorized person could result in your iCare being unsafe for you and your patients Periodical validation is limited exclusively to professional dental installers who have bee...

Страница 11: ...disinfecting solution Once the onsite preparation of the handpieces is complete they will should be safely transported to the reprocessing site avoiding injuries contamination and damage 5 1 2 Preclea...

Страница 12: ...Please refer chapter 10 for the list of available coupling adaptors Fig 14 Connection of a turbine Fig 15 Screwing the retaining nut Fig 16 Coupling adaptor It is necessary to use specific iCare air t...

Страница 13: ...aining ring of the adaptor Remove the turbine by pulling it down while keeping the retaining ring in the upright position Removing a turbine with Kavo and Sirona connection from its adaptor Remove sim...

Страница 14: ...ion Lubrication Partial cycle Cleaning Lubrication Short cycle Lubrication only K7 Spray key In conjunction with K2 to K5 this key is used to define if the handpiece has an external spray channel or n...

Страница 15: ...op corresponds to a normal volume Short cycle Large drop corresponds to a specific volume for certain contra angles long cycle for surgical or 1 5 contra angle for example Unlit light indicators 3 ind...

Страница 16: ...external spray instrument but in reality the instrument has an internal spray An instrument that has been installed with external spray without using the button ext W O An instrument which has been in...

Страница 17: ...urther steps Fig 30 Indicators during cycle in progress Caution When the oil drop indicators are switched off the related instruments are not selected Before first use or after a long period of no use...

Страница 18: ...l panel and never use aggressive detergents 6 2 General maintenance guidelines Frequency Number of cycles Operation Reference Section Every week 50 Cleaning the treatment chamber 6 2 1 Every week 50 C...

Страница 19: ...n and n cid solutions are formulated to be disposed of in the sewage disposal Prior to doing so the soiled fluids should be diluted using water in a 4 1 ratio 4 volume of water for 1 volume of soiled...

Страница 20: ...er use other fuses than those described below Voltage 250VAC Rating T 1 6AH 250V CHAPTER 7 INFORMATION FROM WARNING INDICATORS As shown in Fig 34 warning indicators inform in real time about abnormal...

Страница 21: ...ons To ensure the efficiency of the reprocessing performed by iCare and enabling the traceability of the handpieces reprocessing data is recorded internally and uploaded to the USB key refer to sectio...

Страница 22: ...luminated and a Beep signal can be heard the air pressure is lower than 5 bar or higher than 6 bar To restart the unit first check the air pressure at the inlet of the machine It is necessary to have...

Страница 23: ...bar and 6 bar 5 m Not significant IEC 61010 1 2016 EN 61326 1 2012 ISO 15883 1 ISO 15883 5 ISO 10993 5 Class II b class in accordance with rule 15 of annex IX NSK Europe GmbH Elly Beinhorn Strasse 8 D...

Страница 24: ...ectronic equipment WEEE Directive 2002 96 EC for product and accessory disposal Consult operating instructions Caution refer to attached instructions For indoor use only Use by date Number of items in...

Страница 25: ...012 iCare WH Adaptor Not supplied with the device Must be ordered separately Z1127013 iCare BA Adaptor Not supplied with the device Must be ordered separately Z1127014 4 6 mm diameter tube Supplied wi...

Страница 26: ...r quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 0 Ut during 10ms 0 Ut during...

Страница 27: ...iCare iCare is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of iCare can help prevent electromagnetic interference by maintaining a mini...

Страница 28: ...al Spain SA C M dena 43 El Soho Eur polis 28232 Las Rozas Madrid Spain Tel 34 91 626 61 28 Fax 34 91 626 61 32 info nsk spain es http www nsk spain es NSK United Kingdom Office 4 Gateway 1000 Arlingto...

Отзывы: