CHAPTER 11: SPECIFICATIONS 

11.1 Technical Specifications 

I-Care+ Device 

Electrical data: 

Input voltage: 
Frequency: 
Power: 
Fuse: 

Environmental data: 

Use temperature: 
Humidity rate for use: 
Maximum altitude for use: 
Atmospheric pressure range for use: 
Transportation and storage temperature: 
Humidity for transportation and storage: 
Atmospheric pressure range for transportation: 

Mechanical data: 

Overall dimension: 
Required space for installation: 
Transportation weight: 
Maximum weight, fully loaded, in conditions of use: 
Maximum noise level: 
Working pressure:  
Air filtration: 
Maximum heat output: 

Compliance: 

Electrical safety: 
Electromagnetic Compatibility: 
Performance: 
Cleaning: 
Cytotoxicity: 

Classification: 

Following MDD: 

Other: 

Manufacturer: 
 
Additional features: 

 
85 to 264 VAC 
47 to 63 Hz 
50 VA 
T 1,6AH 250V 
 
At room temperature 22°C to 40°C 
Max 80% @ 31°C, linear decrease to 50% @ 40°C 
2000m 
700-1060 hPa 
0 to 50°C 
10-80% 
500-1060 hPa 
 
H: 405 mm / W: 355 mm / D: 400 mm 
H: 455 mm / W: 455 mm /D: 450 mm 
14kg 
16kg 
< 60 dBA 
>5 bar and <6 bar 
5µm 
Not significant 
 
IEC 61010-1: 2016 
EN 61326-1: 2012 
ISO 15883-1 
ISO 15883-5 
ISO 10993-5 
 
Class II b class in accordance with rule 15 of annex IX 
 
NSK Europe GmbH, Elly-Beinhorn Strasse 8 
D - 65760, Eschborn, Germany 
Fully micro-processor driven and controlled 

Consumables 

n-clean and n-cid: 

Packaging: 
Transportation and storage temperature: 
Biological performance for n-clean 

(1)

 

 
Biological performance for n-cid 

(1)

 

(2)

  

 
Notes 

(*)

 

(1)    

The products must be used only at room temperature 

(18°C / 25°C) 

(2)    

Test criteria: Clean conditions, room temperature 

(2)     

Spectrum activity in 5 minutes after pre-cleaning with 
n-clean 

 
500 ml, PEHD bottle to be pierced 
Transportation: -20°C/+40°C – Storage: -10 to 30°C 
Cleaning effect: Practice test with test soil according to EN 
15883-5 
Bactericidal: EN 13727, EN 13697 
Fungicidal: EN 13624, EN 13697, EN 17387 
Tuberculocidal: EN 14348, EN 13697, 

Mycobacterium 

terrae

Mycobactericidal: EN 14348, EN 17387, 

Mycobacterium 

avium

Virucidal: EN 14476 against enveloped viruses, e. g. HBV, 
HCV, HIV and influenza as well as non enveloped viruses, e. 
g. adeno-, noro-, polio- and polyomaviruses (SV40) 
Virucidal: EN 16777 against enveloped viruses, e. g. HBV, 
HCV, HIV and influenza as well as non-enveloped viruses, e. 
g. adeno- and noroviruses 

 

Содержание iCare+

Страница 1: ...RD 160 iCare OPM ENG V16 04 27 2020...

Страница 2: ...ion of dynamic dental instruments iCare allows the treatment of the inner and outer surfaces of different instruments contra angles turbines etc Complete treatment in less than 17 minutes Up to 4 inst...

Страница 3: ...njury or damage to the device if the safety instructions are not correctly followed Information Text preceded by this symbol contains useful information for using your iCare device 2 2 General safety...

Страница 4: ...mage the device 2 3 Safety labels Adhesive safety labels have been placed on the iCare in important locations These include storage instructions for the bottles the device and information about the de...

Страница 5: ...is a smell of burning switch off the iCare immediately and contact your retailer Do not use the iCare with flammable gases CHAPTER 3 DESCRIPTION OF THE DEVICE 3 1 Items requested for operation Picture...

Страница 6: ...n cid bottle E Air filter Pressure regulator J USB port for recording data Fig 1 Front view Fig 2 Rear view Fig 3 Operating fluids Fig 4 USB port 3 3 Description of the front panel Table 4 1 Start St...

Страница 7: ...ause moisture ingress The device should be placed in a properly ventilated room The device should be kept away from any source of heat 4 3 Installation 4 3 1 Connection to the air system Insert the ai...

Страница 8: ...checked by a certified electrician A fluctuation in the mains voltage or an electromagnetic field which does not comply with the standard may disrupt the operation of the device 4 4 Consumables 4 4 1...

Страница 9: ...with n cid rinse cautiously with water for several minutes Remove contact lenses if present and easy to do Continue rinsing If eye irritation persists seek medical advice attention In case of contact...

Страница 10: ...dation performed by an unauthorized person could result in your iCare being unsafe for you and your patients Periodical validation is limited exclusively to professional dental installers who have bee...

Страница 11: ...disinfecting solution Once the onsite preparation of the handpieces is complete they will should be safely transported to the reprocessing site avoiding injuries contamination and damage 5 1 2 Preclea...

Страница 12: ...Please refer chapter 10 for the list of available coupling adaptors Fig 14 Connection of a turbine Fig 15 Screwing the retaining nut Fig 16 Coupling adaptor It is necessary to use specific iCare air t...

Страница 13: ...aining ring of the adaptor Remove the turbine by pulling it down while keeping the retaining ring in the upright position Removing a turbine with Kavo and Sirona connection from its adaptor Remove sim...

Страница 14: ...ion Lubrication Partial cycle Cleaning Lubrication Short cycle Lubrication only K7 Spray key In conjunction with K2 to K5 this key is used to define if the handpiece has an external spray channel or n...

Страница 15: ...op corresponds to a normal volume Short cycle Large drop corresponds to a specific volume for certain contra angles long cycle for surgical or 1 5 contra angle for example Unlit light indicators 3 ind...

Страница 16: ...external spray instrument but in reality the instrument has an internal spray An instrument that has been installed with external spray without using the button ext W O An instrument which has been in...

Страница 17: ...urther steps Fig 30 Indicators during cycle in progress Caution When the oil drop indicators are switched off the related instruments are not selected Before first use or after a long period of no use...

Страница 18: ...l panel and never use aggressive detergents 6 2 General maintenance guidelines Frequency Number of cycles Operation Reference Section Every week 50 Cleaning the treatment chamber 6 2 1 Every week 50 C...

Страница 19: ...n and n cid solutions are formulated to be disposed of in the sewage disposal Prior to doing so the soiled fluids should be diluted using water in a 4 1 ratio 4 volume of water for 1 volume of soiled...

Страница 20: ...er use other fuses than those described below Voltage 250VAC Rating T 1 6AH 250V CHAPTER 7 INFORMATION FROM WARNING INDICATORS As shown in Fig 34 warning indicators inform in real time about abnormal...

Страница 21: ...ons To ensure the efficiency of the reprocessing performed by iCare and enabling the traceability of the handpieces reprocessing data is recorded internally and uploaded to the USB key refer to sectio...

Страница 22: ...luminated and a Beep signal can be heard the air pressure is lower than 5 bar or higher than 6 bar To restart the unit first check the air pressure at the inlet of the machine It is necessary to have...

Страница 23: ...bar and 6 bar 5 m Not significant IEC 61010 1 2016 EN 61326 1 2012 ISO 15883 1 ISO 15883 5 ISO 10993 5 Class II b class in accordance with rule 15 of annex IX NSK Europe GmbH Elly Beinhorn Strasse 8 D...

Страница 24: ...ectronic equipment WEEE Directive 2002 96 EC for product and accessory disposal Consult operating instructions Caution refer to attached instructions For indoor use only Use by date Number of items in...

Страница 25: ...012 iCare WH Adaptor Not supplied with the device Must be ordered separately Z1127013 iCare BA Adaptor Not supplied with the device Must be ordered separately Z1127014 4 6 mm diameter tube Supplied wi...

Страница 26: ...r quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 0 Ut during 10ms 0 Ut during...

Страница 27: ...iCare iCare is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of iCare can help prevent electromagnetic interference by maintaining a mini...

Страница 28: ...al Spain SA C M dena 43 El Soho Eur polis 28232 Las Rozas Madrid Spain Tel 34 91 626 61 28 Fax 34 91 626 61 32 info nsk spain es http www nsk spain es NSK United Kingdom Office 4 Gateway 1000 Arlingto...

Отзывы: