Troubleshooting
7.4.12 Needle Heater
Table 7-12 Needle Heater Troubleshooting
Symptom
Possible Cause
Recovery
No needle heat
Set Point (SP) value below Present
Value (PV) or the heater is in an
idle state.
Change the SP value within the Heater Controls
Window. See
Disconnected heater.
Verify the needle heater cable connections are
securely attached to the correct ports.
Failed DHC power.
Verify DHC indicator lights for 3.3V, -12V, and
12V are illuminated.
Failed heater.
Verify the needle heater electrical parameters.
The resistance between pins 4 and 5 for the
heating element should be 100 ohms +/- 5 ohms.
The resistance of the RTD sensor between pins
6 and 7 should be 108 ohms, +/- 5 ohms.
7.4.13 Tactile Sensor
Table 7-13 Tactile Sensor Troubleshooting
Symptom
Possible Cause
Recovery
During setup routine,
error indicates that the
tactile sensor is already
engaged
Tactile sensor cap is stuck in the
down position.
Visually inspect the sensor cap for fluid buildup.
If there is fluid, clean as specified in
If there is no fluid, call a
service technician.
The tactile sensor signal can be tested from the
Fluidmove
Tools > IO
menu. Select IO and
inputs. Locate the tactile cap signal display and
monitor its state as the cap is depressed and
released. If the state does not change, the
precision switch may have failed.
Needle-to-height-sensor
offset is inaccurate
Fluid build-up on probe tip or tactile
sensor cap may have caused
inaccurate height sensor
calibration.
1. Clean probe tip/tactile sensor cap. See
2. Perform a “Valve Offsets” routine in
Fluidmove. For assistance, refer to the
Fluidmove User Guide
or Online Help.
Содержание Spectrum II S2-900 C Series
Страница 2: ......
Страница 34: ...1 20 Introduction 1 11 7 Rear View Figure 1 8A S2 900 Rear View 30A Power Manager shown 1 2 3 4 5 7 6...
Страница 36: ...1 22 Introduction 1 11 8 Rear View Open Figure 1 9A S2 900 Rear View Open 1 2 3 4...
Страница 40: ......
Страница 124: ......
Страница 157: ...Calibration and Adjustment 5 33 Figure 5 38 Dispense Lift Table Controls...
Страница 159: ...Calibration and Adjustment 5 35 Figure 5 40 Beacon Configuration Window...
Страница 186: ......
Страница 197: ...Troubleshooting 7 11 Figure 7 1 Scale Setup Menu Figure 7 2 Scale Error Message...
Страница 200: ......
Страница 215: ...Parts Replacement 8 15 Figure 8 11 Replacing a Panel Mounted Fuse...
Страница 218: ......
Страница 223: ...Specifications 9 5 Top View mm Back View mm Figure 9 2 S2 900 System Dimensions 2 of 2...
Страница 225: ...Specifications 9 7 Top View mm Back View mm Figure 9 4 S2 9XXC System Dimensions 2 of 2...
Страница 228: ...A 2 Appendix A Material Safety Information A 2 1 Multemps PS NO 2...
Страница 229: ...Appendix A Material Safety Information A 3...
Страница 230: ...A 4 Appendix A Material Safety Information...
Страница 231: ...Appendix A Material Safety Information A 5...
Страница 232: ...A 6 Appendix A Material Safety Information...
Страница 233: ...Appendix A Material Safety Information A 7...
Страница 234: ...A 8 Appendix A Material Safety Information A 2 2 Moly Graph Extreme Pressure Multi Purpose Grease...
Страница 235: ...Appendix A Material Safety Information A 9...
Страница 236: ...A 10 Appendix A Material Safety Information...
Страница 237: ...Appendix A Material Safety Information A 11...
Страница 238: ...A 12 Appendix A Material Safety Information...
Страница 239: ...Appendix A Material Safety Information A 13...
Страница 240: ...A 14 Appendix A Material Safety Information...
Страница 246: ......
Страница 252: ......
Страница 258: ......