6. Maintenance
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
58
No.600-2111-09
Tube replacement of drainage pump
Pump tubing is recommended to replace the 400 hours in luck or about one year. Please
note that the deterioration of the tube can cause failure of the pump.
(1) Turn off the power supply of the device and disconnect the tube from the pump.
(2) Remove the front cover of the blue pump.
(3) To return to the original state pump.
Please apply the silicon