6. Maintenance
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
51
No.600-2111-09
Cleaning the fluorescence cell
(1) Remove L joint on the right of a fluorescence cell.
(2) Remove four screws of the right side, and remove the board.
(3) Remove two front screws of right side.
(4) Remove four front screws and a front board.
(5) Wipe the inside and the transparent field of a cell with the swab, which attached
ethanol.
(6) Place a fluorescence cell so that the side on which the film is stuck becomes front
side and the bottom, and insert packing over left-hand side.
* Be careful not to attach a fingerprint etc. to a transparent area in the case of work.
(7) Attach packing, a joint and right side board, and tighten screws lightly.
(8) Attach a front board and tighten left-side screws.
(9) Tighten the screws of the front and a right side.
(10) Tighten the screws of the front and a right side little by little equally.