5. Measurement
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
26
No.600-2111-09
5.4. Preparation for measurement
Supply of Career Gas
Open the main cock of an argon gas cylinder and an air gas cylinder. And adjust two pressure
regulators of the PE-1 back to 0.2Mpa.
Open the main cock of supply of argon gas and
air gas. (0.3 - 0.5KPa)
1. Cancel the lock of a pressure regulator. (The
knob of that to the down side.)
2. Pull the knob to release the lock as described.
3. Turn the knob and adjust a pressure to 0.2Mpa.
* If a pressure is turned to the right, it will become
high, and if it turns to the left, it will become low.
4. Push up and lock a knob after adjustment.