6. Maintenance
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
46
No.600-2111-09
(4) Press the
↓
key to select [Injectors Number 1].
(5) Press the
↓
key to move to [Injector ALIGN.].
Press the
→
key to select [Injector ALIGN.].
The syringe moves to the preset injection position.
(6) While using the
←
and
→
keys to adjust the horizontal position, use the
↓
key to
lower the syringe to the sample injection section
(7) Press the ENTER key to determine the position.
(8) Press the
↓
key to lower the syringe to a position just before the syringe is lifted.
(Approximately -43 mm)
(9) Press the ENTER key to determine the syringe injection depth.
(10) Press the SET key to return to the title screen.