5. Measurement
AUTOMATIC PETRO PYROLYSIS MERCURY ANALYZER – MERCURY/PE-1000
31
No.600-2111-09
STD value and Sample Name Registration
Input the standard value into "STD" table and the sample name into "SMP" table.
STD value
(1) Click the cell of "STD" for opening the “STD
INPUT” window.
(2) Set up unit. Click “ppm” or “ppb” button.
(3) Input STD concentration and volume.
* Input Gas Temperature and Volume for RH
mode.
Sample Name
Click the cell of “NAME” and Input a
sample name.
*SVOL is inputted automatically, when
HT-300A is used.
(Refer to 7.1.4Table window)
Measurement of heavy oil
The high clay oil can not sample. Was diluted it to 5 times with toluene, use CRUDE
OIL method. You should use that toluene was low mercury concentration.
If you use a unit of weight / weight, density of the mixture is required.
weight / weight
weight / volume