Chapter 2
Microscopy Operations
40
Ep
i-fluo
resce
n
ce
Micro
sco
p
y
Preparation Focusing
Microscopy
Epi-fluorescence
Microscopy Procedure
1
2
3
4
5
6
7
8
9 10 11 12 13 14
■
■
■
■
■
Viewing the Specimen
12
View the specimen.
To observe another part of the sample, rotate the stage
knobs to find the position to observe. If the sample is not
in focus, rotate the focus knobs to adjust the focus.
To observe under a different magnification, repeat steps
6 and later with another objective.
Diascopic image in fluorescence
observation
For fluorescence observations, press the
microscope's dia-illumination ON/OFF switch to
“OFF” to make the diascopic image disappear.
Bright ambient lights will make it difficult to view the
image. Nikon recommends keeping the room dark
during fluorescence observations.
NI-FA FL/DIC Analyzer Slider/Dummy
Slider
In the epi-fluorescence microscopy, the FL/DIC
analyzer slider inserted in the epi-fluorescence cube
turret is not used. When the analyzer slider is
inserted in the epi-fluorescence cube turret, pull it out
to the first click position to remove the analyzer from
the optical path. When a dummy slider instead of the
analyzer slider is inserted, do not fully insert it. Pull it
out one step. If it is fully inserted, the periphery of
field of view may be dark.
13
Turn off the mercury lamp.
See your illuminator's manual for details.
INTENSILIGHT
C-HGFIE
LAMP
POWER
RUN TIME
hrs.
Turning off the mercury lamp
POWER indicator (green)
OFF
Power switch OFF
Содержание ECLIPSE Ni-E
Страница 2: ......
Страница 24: ...Chapter 1 Components 10 Components ...
Страница 29: ...Chapter 2 Microscopy Operations 15 Before Microscopy Operation Flowchart ...
Страница 30: ...Chapter 2 Microscopy Operations 16 Before Microscopy Operation Flowchart ...
Страница 31: ...Chapter 2 Microscopy Operations 17 Before Microscopy Operation Flowchart ...