Chapter 2
Microscopy Operations
27
Brig
h
t/Dark-field
Micro
sco
p
y
Preparation
Focus and Optical
System Adjustment
Micro-
scopy
Bright/Dark-field
Microscopy
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16 17 18 19
13
Bring a desired objective into the optical path.
Press the OBJ CW or CCW button to bring the desired
objective into the optical path.
Oil immersion
When using an oil immersion type objective, apply Nikon
specified non-fluorescence immersion oil between the
specimen and the objective. (See Chapter 3 “14
Oil/Water Immersion”.)
Never apply oil between the universal condenser top
lens and the specimen. The universal condenser is
designed for dry use.
Note that an oil/water immersion type objective cannot
be used for the motorized universal condenser.
Motorized nosepiece toggling
Two objectives can be alternated by specifying an
objective pair. Objectives can be switched using the
[OBJ CW/CCW] button on the microscope. (See
Chapter 3 “11.1 Motorized Nosepiece Operation”.)
Bringing desired objective into optical path
P A
_
4 0 x
_ _ _ _
Z :
_ _ _ _ _
0 . 0 0 0 u m
O P E N
_ _
A S 3 0 . 0 m m
_ _
F S 3 0 . 0 m m
[NOSEPIECE] button, sub screen: Desired objective
Содержание ECLIPSE Ni-E
Страница 2: ......
Страница 24: ...Chapter 1 Components 10 Components ...
Страница 29: ...Chapter 2 Microscopy Operations 15 Before Microscopy Operation Flowchart ...
Страница 30: ...Chapter 2 Microscopy Operations 16 Before Microscopy Operation Flowchart ...
Страница 31: ...Chapter 2 Microscopy Operations 17 Before Microscopy Operation Flowchart ...