18
Phase Contrast Microscopy
Phase contrast microscopy is suitable for observation of clear and colorless specimens, undyed or lightly colored
specimens, decolored specimens, and ultrathin slices for electron microscopes. The phase contrast method is not suitable
for graded or hard dyed specimens.
Optical elements necessary for the microscopy is a Ph objective (phase plate) and a PH module (Ph annular diaphragm)
on the condenser. For Ni-E, use a universal condenser for which attaching a PH module is available (motorized or
manual).
18.1
Tips for Phase Contrast Microscopy
A phase contrast image appears differently depending on the phase contrast or shape of the specimen, and the properties
of the objective. Note the following when you prepare the specimen or select a Ph objective.
Select a specimen for which the center of the PH module is not misaligned.
The center of the PH module will be misaligned when observing a specimen that causes scattered light or generates a
lens or prism effect. In particular, with live and thick specimens, oversized specimens, and specimens using a microplate,
the center will be misaligned due to the lens or prism effect, making the specimen difficult to observe.
Ph objective and specimen
Ph objectives include “achromatic”, “plan achromatic”, “plan flour”, and “plan apochromatic” objectives depending on how
much the chroma aberration and field curvature are adjusted. These lenses also are further subdivided into several types
depending on the properties of the internal phase plate. For favorable microscopy results, the phase contrast amount of
the specimen must match the properties of the phase plate. See the table below for the use properties of Ph objectives.
When using a dark contrast Ph objective, make sure that the phase contrast of the specimen does not exceed the allowed
phase contrast amount (latitude). If the phase contrast amount of the specimen is greater than the allowed phase contrast
amount, observation is not possible as the image will be illuminated brighter than the background.
You can increase or decrease the phase contrast by the thickness of the specimen, and the refraction index of the
mounting agent or culture solution when preparing a phase contrast specimen.
A specimen with weak contrast under a DLL objective may yield better result under a DM objective.
Use Properties of Ph Objectives
Ph contrast
objective
Appearance
Contrast
Latitude
Usage example
DLL
DL
Intermediate
(with broader
usage)
Phase contrast
and absorbing
object
(chromosome) in
low and
intermediate
latitude
Bacteria's spore, general
live bacteria, slightly thick
specimen, bacteria, dyed
specimen, egg, fat particle,
crystalline, etc.
Dark
contrast
DM
Generally, an object with
larger phase contrast
appears darker.
Therefore, the image is
shown in black in a
relatively brighter field of
view, similar to the one
observed with
bright-field microscopy.
Suitable for
detailed
observation
mainly using
micro
contrast.
High
(with
relatively
narrower
usage)
Transparent
object in low
latitude
Bacteria and protozoal
flagellum, fibrin basic fiber,
fine granule,
mounting-agent-selective
slice, ultrathin slice, etc.
Bright
contrast
BM
Generally, an object with
larger phase contrast
appears brighter.
Therefore, the image is
shown brighter in a
relatively darker field of
view, similar to the one
observed with dark-field
microscopy.
Suitable for morphology,
detection, and calculation of
fine fiber and granule mainly
using macro contrast.
Almost all areas
Bacteria and protozoal
flagellum, fibrin basic fiber,
fine granule, blood cell
calculation, etc.
Содержание ECLIPSE Ni-E
Страница 2: ......
Страница 24: ...Chapter 1 Components 10 Components ...
Страница 29: ...Chapter 2 Microscopy Operations 15 Before Microscopy Operation Flowchart ...
Страница 30: ...Chapter 2 Microscopy Operations 16 Before Microscopy Operation Flowchart ...
Страница 31: ...Chapter 2 Microscopy Operations 17 Before Microscopy Operation Flowchart ...