27
5.3 Trace Data is not Normal
5.3.1 Frame trace data is not normal
Is frame tracing conducted?
No
5.3.1 Frame trace data is not normal (p27)
Yes
Is pattern tracing conducted?
No
5.3.2 Pattern trace data is not normal (p28)
Is the attachment of the stylus (40601-7910) loose?
Yes
Attach the stylus (40601-7910) securely
No
Is the stylus (40601-7910) worn out?
Yes
Replace the stylus (40601-7910)
No
Conduct the standard frame calibration
Conduct full calibration
Replace the tracer ASSY (40601-0010)