20
5.1.3 Error occurs
5.1.3.1 ERROR2: No trace data
Conduct tracing again.
5.1.3.2 ERROR17: Stylus came off (lower)
The LT-980 is not equipped with error display function.
Use the Error History display function of Tracer Utility to confirm error contents
ERROR20, ERROR51, and ERROR1000 and later occur when the tracer ASSY (40601-
0010) is defective.
Replace the tracer ASSY (40601-0010)
.
When repairing the tracer ASSY (40601-0010), refer to the service manual for the tracer
ASSY (LETR-1).
Is the trace position beyond the traceable range?
Yes
Set the frame in the center and conduct tracing
again.
No
Conduct goggle type frame tracing.
Is frame tracing conducted although a pattern is
set?
Yes
Replace the tracer ASSY (40601-0010)
No
Check the calibration data (refer to the service
manual for the LETR-1).
Is the calibration data normal (refer to the service
manual for the LETR-1)?
No
Conduct full calibration
.
Yes
Replace the tracer ASSY (40601-0010)
Содержание LT-980
Страница 1: ...Satellite Tracer LT 980 SERVICE MANUAL October 2012 Pages in total 82 LLT9A RDA002A...
Страница 2: ......
Страница 6: ...LLT9A RDA002A 6 This page is intentionally left blank...
Страница 8: ...LLT9A RDA002A 8 This page is intentionally left blank...
Страница 12: ...LLT9A RDA002A 12 This page is intentionally left blank...
Страница 14: ...LLT9A RDA002A 14 This page is intentionally left blank...
Страница 32: ...LLT9A RDA002A 32 This page is intentionally left blank...
Страница 44: ...LLT9A RDA002A 44 This page is intentionally left blank...
Страница 57: ...LLT9A RDA002A 57 5 Remove the standard frame 40601 7500 40601 7500 Circumference...
Страница 69: ...LLT9A RDA002A 69 9 3 Wiring Diagram...
Страница 70: ...LLT9A RDA002A 70 9 4 Connector Cable 1 44801 CA01 2 44801 CA02 3 44801 CA03 4 44801 CA04...
Страница 80: ...LLT9A RDA002A 80 This page is intentionally left blank...
Страница 81: ......