56
8.2.2 Standard frame calibration
1 . Turn on the power switch of the LT-980.
When initialization is complete, the indica-
tor above
illuminates.
2 . Activate Calibration mode.
1 ) While pressing
, press
.
A short beep sounds.
2 ) Release
and .
The indicator above
blinks.
3 . Set the standard frame (40601-7500) to
the tracer.
4 . Press .
Calibration starts.
When calibration is complete, a short
beep sounds and the indicator above
illuminates.
The standard frame is released.
Use the standard frame (40601-7500) included in the
standard accessories
.
Caution
With the side on which the cir-
cumference is indicated facing
up, set the standard frame
(40601-7500) so that the rim
clips of the upper and lower
sliders are inserted into the
groove of the frame.
Standard frame
40601-7500
Circumference
Содержание LT-980
Страница 1: ...Satellite Tracer LT 980 SERVICE MANUAL October 2012 Pages in total 82 LLT9A RDA002A...
Страница 2: ......
Страница 6: ...LLT9A RDA002A 6 This page is intentionally left blank...
Страница 8: ...LLT9A RDA002A 8 This page is intentionally left blank...
Страница 12: ...LLT9A RDA002A 12 This page is intentionally left blank...
Страница 14: ...LLT9A RDA002A 14 This page is intentionally left blank...
Страница 32: ...LLT9A RDA002A 32 This page is intentionally left blank...
Страница 44: ...LLT9A RDA002A 44 This page is intentionally left blank...
Страница 57: ...LLT9A RDA002A 57 5 Remove the standard frame 40601 7500 40601 7500 Circumference...
Страница 69: ...LLT9A RDA002A 69 9 3 Wiring Diagram...
Страница 70: ...LLT9A RDA002A 70 9 4 Connector Cable 1 44801 CA01 2 44801 CA02 3 44801 CA03 4 44801 CA04...
Страница 80: ...LLT9A RDA002A 80 This page is intentionally left blank...
Страница 81: ......