25
5.1.3.14 ERROR53: Shape calculation error
5.1.3.15 ERROR240: Calibration jig error
Conduct full calibration with a proper jig
5.1.3.16 ERROR1001 and later
Perform the procedure as in “7.1.1 Tracer ASSY (40601-0010)” (p33).
Does a frame, pattern, or dummy lens have a sharp
change in its shape?
Yes
Tracing is not possible.
No
Set the frame in the center and conduct tracing
again.
Replace the tracer ASSY (40601-0010)
ERROR1000 and later occurs when the tracer ASSY (40601-0010) is defective.
When repairing the tracer ASSY (40601-0010), refer to the service manual for the tracer
ASSY (LETR-1).