18
5.1.2 Initialization is not conducted properly
The indicator above
blinks during initialization.
When initialization is complete properly, the indicator above
illuminates.
Does the indicator above the button illuminate?
No
5.1.2.1 Indicator does not illuminate (p19)
Yes
Turn off the power switch.
Remove the top cover ASSY (44801-2000)
.
Are there any breaks in the cable ASSY (44801-
CA04)
Yes
Replace the cable ASSY (44801-CA04).
No
Are the DIP SWs on the main board (40601-BA01)
set properly
No
Set the DIP SWs on the main board (40601-BA01)
properly.
Yes
Replace the tracer ASSY (40601-0010)