MultiPro 2000 User Manual
23
Patient Safety Precaution:
SINCE THE TEST METHOD INJECTS POTENTIALLY HAZARDOUS
CURRENT LEVELS INTO THE ELECTROCARDIOGRAPH AND THE
RELATED POWER SYSTEM, DO NOT CONDUCT TESTS IN AN
OCCUPIED PATIENT LOCATION OR WHILE THE PATIENT IS
CONNECTED TO A RELATED POWER SYSTEM BRANCH CIRCUIT.
This test verifies that the isolation system on the device
under test is operating properly.
The following block diagram shows the electrical
connections between The Safety Analyzer and the device
under test.
CAUTION:
THE EKG ISOLATION VOLTAGE IS EQUAL OR APPROXIMATELY
10% HEIGHER THAN THE AC MAINS AND ALTHOUGH IT IS
CURRENT LIMITED TO 1 mA, IT IS POTENTIALLY LETHAL.