MultiPro 2000 User Manual
14
Key. The display will change to the “INTERNAL
MEASUREMENT “Menu Press F1 Key (“VOL”) and
the line voltage will be displayed.
Press “ESC” to go back to the previous Menu
(“INTERNAL MEASUREMENT “Menu).
(“CUR”):- Device Current Measurement.
This function measure the current consumed by the
DUT. To measure the current, press F2 (“CUR”) from
the “INTERNAL MEASUREMENT“ Menu. Press
“READ” to read the device current.
If the out let is off, normal polarity will be activated
and the unit (DUT) will turn on.
Press “ESC” to go back to the previous Menu
(“INTERNAL MEASUREMENT “Menu).
(“POL”) Polarity Check of the Inlet Power.
From the “INTERNAL MEASUREMENT “Menu select
“POL” by pressing F3. The polarity of the incoming
AC power will be checked by measuring the voltage
between HOT and Neutral, HOT and Earth ground as
well as and Neutral and Earth ground. While
checking polarity the display will indicate
“Checking………” for a brief period and the polarity
will then be displayed as “Normal Polarity” or
“Reverse Polarity” depending on the result.
If an “Error” message appears either the outlet the