Physical Dimensions:
Size: 10.25” X 6.25” X 2.5”
Weight: 4 lbs.
Temperature:
Operating range: 15
°
C - 35
°
C (59
°
F -95
°
F)
Storage Temperature: 0
°
C - 50
°
C (32
°
F -122
°
F)
Power:
110 Volt Units (PN: 560-110): 90 V AC to 130 V, 50
Hz to 60 Hz, 20 A maximum.
220 Volt Units (PN: 560-220): 180 VAC to 260 V AC
to 130 V, 50 Hz to 60 Hz, 16 A Maximum.
Simulator:
NSR:
(Normal Sinus)
Rate: 30, 60, 70, 80, 90, 100, 120, 150, 180,
210, 240, 270 and 300 BPM.
Accuracy: +1% (+ /-1 BPM).
Amplitude: user selectable from 0.1 to 4.5mV.
Accuracy: + 2 % (+0.05mV), peak to peak.
Performance:
(SINE, SQUARE, TRIANGLE, PULSE)
Amplitude: Same as NSR.
Accuracy: + 5 % of the reading (+ 0.1mV).
Rates: 0.1 to 0.5 Hz, 1 to 9 Hz, in 1 Hz
Increments, 10 to 60 Hz in 5 Hz increments.
Accuracy: +5%