MultiPro 2000 User Manual
10
The test receptacle used in the Multi-Pro is a Hospital
Grade 20 Amp (NEMA-5-20R). It will accept all standard
line cords as well as horizontally oriented neutral prongs
on 20 Amp line cords.
Various International power outlet configurations are
available. Check with factory for the specific Country.
Power to the test receptacle is in the off mode. Manual
selection of
normal
or
reverse
polarity must be selected
using the keypad to energize the test receptacle during
leakage measurement functions.
Menu Functions and Operation
The operation of the MultiPro 2000 is Menu driven. It is
simple and easy to use following the interactive Menus
with soft keys (F1,F2, F3, F4, F5, F6) and dedicated key
functions such as “Simulator” “Case Leakage”, Leads
Leakage”, Resistance Measurement”.
Main Menu
Following are the abbreviation and description of each
“Main Menu” functions.
•
“
INT
”- Internal measurements.
Additional menu selections in this category are:-
“
VO
L”- Input Voltage,
“
CUR
”- Device current
“
POL
”- Polarity check of the AC input.
Refer to “Measurement Section for details
.