MultiPro 2000 User Manual
25
Simulator Function:
The Simulator function menu is activated by the
“Simulator” key. The following is the screen shot of the
Simulator Menu (
Screen Shot 6
). The following simulator
selections are the available.
•
NSR – Normal Sinus Rhythm.
o
Rates:
Rates from 30 to 300 BPM.
o
Amplitudes:
0.1 to 4.5 mV.
•
Performance Waveform.
o
SINE, TRIANGLE, SQUARE, PULSE
o
Rates
: 0.1 to 60 Hz
o
Amplitudes:
0.1 to 4.5 mV.
Connect the ECG leads to the respective ECG Snap posts.
If the DUT is off, press the Normal Polarity to turn on the
unit.
Select the desired waveform using F1, F2, F3, F4 or F5,
from the Simulator Menu.
To change the rate, press F1 and select the desired Rate
using F1 to F6 keys. The rate selected will be saved in the
memory until it is changed again. Press “ESC” to go back
to the previous Menu.
Screen Shot 6