Specifications
80
Table 13: Tone Definition
Audible Indicator
Parameter
Value
Alarm Silence
Reminder
Volume level
Not changeable
Pitch (±30 Hz)
784 Hz
Pulse width (±20 msec)
150 msec
Pulse repetition interval
(±20 msec) (double burst)
150 msec
Repetitions
3
Confirmation of
Button Pressed
Volume level
Not changeable
Pitch (±30 Hz)
784 Hz
Pulse width (±20 msec)
150 msec
Pulse repetition interval
(±20 msec) (double burst)
150 msec
Repetitions
3
High Priority Alarm
Volume level
Adjustable alarm
volume
Pitch (±30 Hz)
932 Hz
Pulse width (±20 msec)
Nellcor = 255
msec
IAW IEC 60601-
1-8 = 120 msec
IAW EN 475 =
150 msec
Pulse repetition interval
(±20 msec) (double burst)
Nellcor = 320
msec
IAW IEC 60601-
1-8 = 6940 msec
IAW EN 475 =
7500 msec
Repetitions
Continually
Invalid Button Press
Volume level
Not changeable
Pitch (±30 Hz)
180 Hz
Pulse width (±20 msec)
70 msec
Pulse repetition interval
(±20 msec) (double burst)
N/A
Repetitions
1
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......