![Nellcor OXIMAX N-560 Скачать руководство пользователя страница 19](http://html2.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_service-manual_3907265019.webp)
Performance Verification
N-560 Service Manual
9
Figure 3: N-560 Controls
Power-On Self-Test (POST)
CAUTION: If any indicator or display element does not light, or the
speaker does not sound, do not use the N-560. Instead, contact qualified
service personnel, your local Nellcor representative, or Nellcor's
Technical Services Department, 1.800.635.5267.
Note:
The N-560 should complete the POST function within 12 seconds.
1. Turn on the N-560 by pressing the
Power On/Off
button for greater than
0.5 seconds.
2. The N-560 displays/sounds:
1 —
Power On/Off
Button
5 —
Pulse Rate Alarm Limit
Button
2 —
Alarm Silence
Button
6 — SatSeconds
Alarm Limit Button
3 —
Adjust Up Button
7 —
SpO
2
Alarm Limit Button
4 —
Adjust Down Button
Display
Sound
5
(in pulse rate left window)
low priority alarm
6 (in pulse rate center window)
medium priority alarm
0 (in pulse rate right window)
high priority alarm
n (in SpO
2
left window)
n (in SpO
2
center window)
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......