Routine Maintenance
6
2. Perform the electrical safety tests detailed in
Safety Tests
on page 26. If
the N-560 fails these electrical safety tests, refer to
Troubleshooting
on
page 43.
3. Inspect the fuses for proper value and rating (F1 and F2 = 2 amp,
250 volts).
Battery
Nellcor recommends replacing the N-560's battery every two years. When the
N-560 is going to be stored for two months or more, remove the battery prior
to storage. To replace or remove the battery, refer to
Disassembly Guide
on
page 53.
If the N-560 has been stored for more than 30 days, charge the battery as
described in
Battery Charge
on page 8. A fully discharged battery requires
6 hours to receive a full charge. The battery is being charged whenever the
N-560 is plugged into an AC power source.