![Nellcor OXIMAX N-560 Скачать руководство пользователя страница 103](http://html2.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_service-manual_3907265103.webp)
Specifications
N-560 Service Manual
93
Patient Risk Applied Current
Patient Isolation Risk Current
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1,
UL 2601
-
1
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
100 µA
10 µA
Normal
Open
Closed
500 µA
50 µA
Normal
Closed
Open
500 µA
50 µA
Reversed
Closed
Closed
100 µA
10 µA
Reversed
Open
Closed
500 µA
50 µA
Reversed
Closed
Open
500 µA
50 µA
AC Line
Polarity
Neutral Line
Cord
Power Line
Ground
Cord
IEC 60601
-
1,
UL 2601
-
1
AAMI/
ANSI
-
ES1
Normal
Closed
Closed
5 mA
50 µA
Reversed
Closed
Closed
5 mA
50 µA
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......