Audible Alarm Settings and Service Functions
N-560 Service Manual
31
4
-
Data port baud rate
24
2400 bauds per second (bps)
96
9600 bps
192
19200 bps
5
-
EPP Mode
1
Real Time ASCII
2
External equipment communications
6
-
In-sensor Trend Mode
0
Event SpO
2
1
Event SpO
2
+ Pulse Rate
2
Not use in-sensor trend function
7
-
RS-232 and RS-422 Nurse Call polarity (negative -
low, positive - high)
0
Positive on alarm, normally negative
1
Negative on alarm, normally positive
8
-
Time Set - Each display “-XX” (“-” means Selected.
“XX” means Numbers) Use
Adjust Up
/
Adjust
Down
buttons to change numbers. Use
SatSeconds
Alarm Limit button to save selection.
Year
1
ST
display — “-XX” at left numeric display
Month
1
ST
display — “-XX” at right numeric display
Day
2
ND
display — “-XX” at left numeric display
Hour
2
ND
display — “-XX” at right numeric display
Minute
3
RD
display — “-XX” at left numeric display
Second
3
RD
display — “-XX” at right numeric display
9
-
Save current N-560 parameters to institutional
(Power-On default settings)
10
-
Restore N-560 parameters to factory default settings
Table 2: N-560 Menu Selections (Continued)
Menu
Number
Sub-Menu
Number
Function
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......