![Nellcor OXIMAX N-560 Скачать руководство пользователя страница 17](http://html2.mh-extra.com/html/nellcor/oximax-n-560/oximax-n-560_service-manual_3907265017.webp)
N-560 Service Manual
7
P er f or ma n ce Ve r i f i ca t i o n
Introduction
This section discusses the tests used to verify performance following repairs
or during routine maintenance. All tests can be performed without removing
the N-560 cover. All tests except the battery charge and battery performance
tests must be performed as the last operation before the N-560 is returned to
the user.
If the N-560 fails to perform as specified in any test, repairs must be made to
correct the problem before the N-560 is returned to the user.
Equipment Needed
Table 1: Equipment Needed
Equipment
Description
Digital multimeter (DMM)
Fluke Model 87 or equivalent
Durasensor
®
oxygen transducer
(sensor)
DS-100A
O
XI
M
AX
oxygen transducer
MAX-A
Pulse oximeter tester
SRC-MAX (must meet current EN
61326-1, FCC Part 15, and UL 1010-1,
or IEC 61010-1, or CSA 22.2 No.
1010-1 specifications)
Safety analyzer
Must meet current AAMI ESI/1993 &
IEC 60601-1/1998 specifications
Pulse oximetry cable
DOC-10
Data interface cable
EIA-232 cable (optional)
Stopwatch
Manual or electronic
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......