Specifications
N-560 Service Manual
91
Table 18: Recommended Separation Distances Between
Portable and Mobile RF Communication Equipment and the
N-560
The N-560 is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the N-560 can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the N-560
as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
Output Power of
the Transmitter
W
Separation Distance according to Frequency of
Transmitter
m
150 kHz to 80
MHz
80 MHz to 800
MHz
800 MHz to 2.5
GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance
d
in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where
P
is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1:
At 80 MHz to 800 MHz, the separation distance for the higher
frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
Содержание OXIMAX N-560
Страница 1: ...C M Y CM MY CY CMY K Service ai 1 24 06 9 25 48 AM Service ai 1 24 06 9 25 48 AM...
Страница 10: ...List of Tables N 560 Service Manual viii...
Страница 38: ...Performance Verification 28...
Страница 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Страница 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Страница 84: ...Spare Parts 74...
Страница 88: ...Packing for Shipment 78...
Страница 104: ...Specifications 94...
Страница 114: ...Data Port Interface Protocol 104...
Страница 140: ...Technical Discussion 142...
Страница 144: ......
Страница 145: ......